XPiQ inc.
Intelligent Design Quality Product
Medical Switching Power Supply
200 Watts
PM200 Series
•
4000 VAC Isolation
•
100% Burn-in
•
Low Safety Leakage Current
•
Short Circuit Protection
•
Meets EMI Requirements of
Medical Equipment
•
Both Line and Neutral Fused for
European Requirements
Specification
All specifications typical at nominal line, full load and 25°C
Input
Input Voltage
Input Frequency
Input Current
Safety Ground
Leakage Current
• 90 to 130 VAC or 180 to 264 VAC
autoranging
• 47 to 440 Hz
• 4.5A (rms) for 115 VAC
2.6A (rms) for 230 VAC
• 50µA max. at 110 VAC 60 Hz
• 100µA max. at 240 VAC 50 Hz
Output
Total output power: •
Ripple and noise: •
Overvoltage
•
Protection:
Overcurrent
•
Protection:
Temperature
•
Coefficient:
Transient Response: •
Line Regulation
•
15 to 200 watts
2% peak to peak maximum
Provided on output #1 only;
set at 112-132% of its nominal
output voltage
All outputs protected to short
circuit conditions
All outputs ±0.04%/°C
maximum
Maximum excursion of 4% or
better on all models, recovering
to 1% of final value 500us after
a 25% step load change
±0.5% max. at full load
General
Efficiency
•
Hold-up time
•
Inrush Current
•
Withstand Voltage •
Environmental
Operating
•
Temperature
Storage
•
Temperature
Relative Humidity •
MTBF
•
EMI Requirement •
Safety Requirements •
70% minimum at 200 watt output
20 msec minimum at nominal input
60A max at 25°C cold start
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground
0°C to +70°C*
-40°C to +85°C
5% to 95% non-condensing
85,000 hours minimum at full load
at 25°C ambient
Meets the conducted limits of
EN 55011 (CISPR11) Level B
Approved to: (Not for patient
contact)
a) UL2601
b) CSA C22.2 No.601.1-M90 per CUL
c) IEC 601-1/EN60601-1 (per TUV)
157 www.XP-iQ.com
* Derate linearly from 100% load at 50°C to 50% load at 70°C
PH: 508 429.9883
FAX: 800 226.2100
Email: sales@xp-iq.com
Holliston, MA 01746 USA