PA1000 Power Analyzer
Specifications
All specifications are guaranteed unless noted otherwise. All specifications apply to all models unless noted otherwise.
Available measurements
Vrms - Volts RMS
Arms - Amps RMS
WATT - True power
VA - Apparent power
VAR - Reactive power
FREQ - Frequency
PF - Power factor
VPK+ - Volts peak (positive)
VPK- - Volts peak (negative)
APK+ - Amps peak (positive)
APK- - Amps peak (negative)
VDC - Volts DC
ADC - Amps DC
VCF - Voltage crest factor
ACF - Current crest factor
VTHD - Volts Total Harmonic Distortion
VDF - Voltage distortion factor
ATHD - Amps Total Harmonic Distortion
ADF - Current distortion factor
Z - Impedance
R - Resistance
X - Reactance
HR - Integrator time
WHr - Watt Hours
VAHr - VA Hours
VARHr - VAR Hours
AHr - Amp Hours
Vh - Volts harmonics
Ah - Amps harmonics
Voltage and current ranges
Voltage ranges
Current ranges (20 A shunt)
Current ranges (1 A shunt)
1000 Vpeak, 500 Vpeak, 200 Vpeak, 100 Vpeak, 50 Vpeak, 20 Vpeak, 10 Vpeak
100 Apeak, 50 Apeak, 20 Apeak, 10 Apeak, 5 Apeak, 2 Apeak, 1 Apeak, 0.5 Apeak, 0.2 Apeak, 0.1 Apeak
2.0 Apeak, 1.0 Apeak, 0.4 Apeak, 0.2 Apeak, 0.1 Apeak, 0.04 Apeak, 0.02 Apeak, 0.01 Apeak, 0.004 Apeak, 0.002 Apeak
Measurement accuracy - voltage
Voltage accuracy, VRMS
(45 Hz to 850 Hz)
± 0.04% of Reading ± 0.04% of Range ± 0.005 V
Voltage accuracy, VRMS
(10 Hz to 45 Hz,
850 Hz to 1 MHz, typical)
± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± 0.05 V
Voltage accuracy, DC (typical) ± 0.1% of Reading ± 0.1% of Range ± 0.05 V
Effect of common mode
(typical)
100 V, 100 kHz < 500 mV
Measurement accuracy - current
Current accuracy, ARMS
(45 Hz to 850 Hz) 2
± 0.04% of Reading ± 0.04% of Range ± (1.8 µV/Zext)
Current accuracy, ARMS
(10 Hz to 45 Hz,
850 Hz to 1 MHz, typical)
± 0.1% of Reading ± 0.1% of Range ± (0.02*F)% of Reading ± (50 µV/Zext)
Current accuracy, DC (typical) ± 0.1% of Reading ± 0.1% of Range ± (100 µV/Zext)
2 Offset is valid in low bandwidth, with internal shunts, and after a manual zero has been performed. Offset is 10 µV/Zext in high bandwidth, and with external shunt.