1 2
AC/DC Medical Desktop
90 Watts
PCM90 Series
XPiQ inc.
Intelligent Design Quality Product
Active PFC, Meets EN61000-3-2, -3
Low Leakage Current
Y
AR Single, Dual and Triple Outputs
IN
PRELIM2003 Both Line and Neutral Fused
for European Applications
Q4
4000 VAC Isolation
Medical Safety Agency Approvals
Specification
Input
Input Voltage
Input Frequency
Power Factor
Input Current
Inrush Current
Safety Ground
Leakage Current
Output
Output Voltage
Output Power
Minimum Load
Line Regulation
Load Regulation
Ripple and Noise
Transient Response •
Temperature
coefficient:
Hold-up time
Overvoltage
Protection
Overcurrent
Protection:
90-264 VAC
47 to 63 Hz
0.99 at 100 VAC or 0.94 at
240 VAC
1.3 A maximum
15 A max at 115 VAC or
30 A max at 230 VAC cold start
0.3 mA max at 115 VAC, 60 Hz
0.3 mA max at 230 VAC, 50 Hz
General
Efficiency
MTBF
Isolation Voltage
Insulation Resistance •
Size
Weight
83% typical
100,000 hours minimum at full load
at +25 °C ambient
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground
100 Mohm minimum from
output to ground
6.63” x 3.07” x 1.97”
800 grams approx
See Tables
90 Watts
0 minimum load required for single
output models
±1% maximum at full load
±5% maximum, except 3.3 V model
is ±7%
1% pk-pk maximum
Maximum excursion of 4% or
better on all models. Recovers
to 1% of final value within
500 µs after a 25% step load change
All outputs ±0.04%/°C
maximum
16 msec minimum
Provided on output #1 only
112 to 132% recycle input to reset
All outputs 110-150%
with auto recovery
Environmental
Operating
Temperature
Humidity
Storage
Temperature
EMC & Safety
EMI Requirement •
Safety Approvals
0 °C to +70 °C
Derate linearly from 100% load at
+40 °C to 50% load at +70 °C
5 to 95% RH, non-condensing
-40 °C to +85 °C
EN55011 (CISPR11) Class B
conducted and radiated
UL2601-1, CSA C22.2 No 601.1
per cUL EN60601-1 CE Mark,
IEC 60601-1
* Preliminary datasheet - see website for current specs