NJU9214
[ Ex.3 ] In DCV400mV range, when overflow is occurred on the positive voltage side by 401mV input.
Measured value
Format pattern
overflow on the " + " side
DCV, A=+∗∗∗.∗∗m, □[CR]
"∗∗∗.∗∗" shows overflow, "+" shows overflow on the positive side.
[ Ex.4 ] In DCV400mV range, when overflow is occurred on the negative voltage side by -401mV input.
Measured value
Format pattern
overflow on the "-" side
DCV, A=-∗∗∗.∗∗m, □[CR]
"∗∗∗.∗∗" shows overflow, "-" shows overflow on the negative side.
[ Ex.5 ] In DCV400mV range, when the battery life detector warns under the condition of Ex.1.
Measured value
Format pattern
100.01mV
DCV, A=+100.01m, B[CR]
Battery life detector
[ Ex.6 ] In ACA400mA range, when the following current is measured.
Measured value
Format pattern
40.00mA
ACA, A=+040.00m, □[CR]
[ Ex.7 ] In resistance 400kΩ range,
Measured value
Format pattern
100.10kΩ
when the following resistance is measured.
OHM, A=+100.10k, □[CR]
[ Ex.8 ] In frequency 10kHz range, when the following frequency is measured.
Measured value
Format pattern
5.35kHz
FRQ, A=+005.35k, □[CR]
[ Ex.9 ] In tachometer 6000rpm range, the following revolution is measured.
Measured value
Format pattern
2500rpm
rpm, A=+02500□, □[CR]