DTL-IFB-485
Interface Board for Serial-Input Electronic Loads
WinBurn™ DTLx485 Commands: Set Address
The Active X Control for a particular DTL-IFB-485 unit, will
then have its Address set to match the Rotary Switches on
the DTL-IFB-485. Physically, the rotary switches set the
Hex Address for a particular DTL-IFB-485. Example: A123
is equal to 7B in Hex so rotary switch SW2 is set to "7" and
SW3 is set to "B", while the DTLx485 Control is has its Row
set to "7" and its Column set to "B". Refer to Figure15 for
examples of dimensioning the Rows and Columns in code.
Whereas each rotary switch has 16 settings, in the coding
example of Figure 15, arrays have been setup for the rows
and columns.
The desired load current of any DTL-IFB-485 and DTLXA
Series of 100 Watt loads is then entered in the "Set
Current" text box. Figure 15 coding shows how a low and
high value setting can be used, should dynamic loading
and/or power-cycle burn-in applications be desired.
Users must be careful to have provided appropriate
heatsink/cooling to the DTLXA Series 100W electronic load
to avoid catastrophic failure. The DTLXA Series is rated for
only 10 Watts of power dissipation without external
heatsink/cooling. Clicking on the "Set Current" button, will
now update the load with the value entered in the text box.
The DTL-IFB-485 has an isolated on-board 12-bit A/D,
typically used to read the voltage of a power source (DUT)
under test or burn-in. Clicking on the "Read Voltage" button
of the DTLx485 Control, will return the Voltage reading of
the DUT. User's can also poll the DTL-IFB-485's Fault line.
The Fault line goes active whenever the DTLXA Series falls
below its compliance voltage (typically 2.5V, but 0.6 Volts
for model DTL2A-LC). Users desiring to reset the load to
no current, can click on the DTLx485 Control button "Reset
Load".
A "timer" function can be used, should users desire to
perform a dynamic burn-in, switching between high and
low-levels of current. Review the General Declarations
coding for a Form Load on the next page for examples of
this usage.
The nature of the protocol of interfacing through an RS-485
port to communicate to the DTL-IFB-485 boards does not
lend itself to fast updates (although all loads can be
updated at the same time). Although users can control up
to 256 DTL-IFB-485s from a single PC port, they may find
that 128 units on a single RS-485 port, provides
reasonable update times for burn-in applications. Also,
multiple ports on a single PC approach can be used to
extend the # of units beyond the 256 limit of a single COM
port.
Figure 14. OK or Open Port Invalid Message
13