AC/DC Medical Open Frame
65 Watts
PM65 Series
•
XPiQ inc.
Intelligent Design Quality Product
4000 VAC Isolation
•
100% Burn-in
•
Low Safety Leakage Current
•
Wide Input Voltage 85 to 264 VAC
•
Meets EMI Requirements of
Medical Equipment
•
Both Line and Neutral Fused for
European Requirements
Specification
Input
Input Voltage
•
Input Frequency
•
Input Current
•
Safety Ground
•
Leakage Current
•
Output
Total output power: •
Ripple and noise:
•
Overvoltage
•
Protection:
Overcurrent
•
Protection:
Temperature
•
Coefficient:
Transient Response •
Line Regulation
•
85 to 264 VAC
47 to 440 Hz
1.6 A (rms) for 115 VAC
1.0 A (rms) for 230 VAC
50 µA max at 110 VAC 60 Hz
100 µA max at 240 VAC 50 Hz
0 to 65 watts
1% pk-pk maximum
Provided on output #1 only;
set at 112-132% of its nominal
output voltage
All outputs protected to short
circuit conditions
All outputs ±0.04%/°C
maximum
Maximum excursion of 4% or
better on all models, recovering
to 1% of final value 500 µs after
a 25% step load change
±0.5% max at full load
General
Efficiency
•
Hold-up time
•
Inrush Current
•
Withstand Voltage •
Insulation Resistance •
Switching Frequency •
Environmental
Operating
•
Temperature
Storage
•
Temperature
Relative Humidity •
MTBF
•
EMC & Safety
EMI Requirement •
Safety Requirements •
70% minimum at 65 watt output
10 msec minimum at nominal input
15 A at 115 VAC or 30 A at
230 VAC at +25 °C cold start
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground
10 Mohm minimum from
output to ground
32 kHz (±5 kHz)
0 °C to +70 °C, derate linearly from
100% load at +50 °C to 50% load
at +70 °C
-40 °C to +85 °C
5% to 95% non-condensing
100,000 hours minimum at full load
at +25 °C ambient
Meets the conducted limits of
EN 55011 (CISPR11) Level B
Approved to:
(Not for patient contact)
a) UL2601
b) CSA C22.2 No.601.1-M90 per CUL
c) IEC 601-1/EN60601-1 (per TUV)
•
www.XPiQ.com
PH: 508 429.9883
•
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA